Quality control in medical radiology is a process used by a regulatory authority to evaluate the performance of a diagnostic X-ray machine, including acceptable limits and the actions necessary to meet regulatory requirements. This study evaluates the radiation dose outputs of X-ray equipment in diagnostic medical hospitals by measuring mAs linearity, kVp accuracy, and reproducibility as quality control parameters to ensure that radiation doses administered to patients meet standard regulatory requirements. Radiation measuring instruments were used in this work to measure mAs, kVp accuracy, and reproducibility outputs for five (5) medical diagnostic X-ray machines. The detector was placed on the couch at a distance of 100 cm from the collimator to take measurements for each parameter under investigation. Dose rate measurements for each parameter were recorded one at a time. The mathematical equations for mAs linearity, kVp accuracy, and reproducibility were used to compute the linearity coefficient, percentage error, and coefficient of variation. The computed figures for mAs linearity for the five X-ray machines show that four (4) passed and one fell below the allowable limit. Data for kVp accuracy indicate that three of the machines are within the allowable limits and two (2) failed, while the calculated values for reproducibility show that only two (2) machines passed the tests. The study generates empirical quality control data rarely reported in X-ray diagnostic centers. It verifies the linearity coefficient, percentage kVp error, and the coefficient of variation with practical knowledge in radiology studies. Additionally, it provides a systematic mathematical approach to compute mAs linearity, kVp accuracy, and reproducibility for teaching radiology in non-diagnostic laboratory settings.
| Published in | Radiation Science and Technology (Volume 12, Issue 2) |
| DOI | 10.11648/j.rst.20261202.12 |
| Page(s) | 31-43 |
| Creative Commons |
This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited. |
| Copyright |
Copyright © The Author(s), 2026. Published by Science Publishing Group |
kVp Accuracy, mAs Linearity, Quality Control, Radiology, Reproducibility, X-ray
No | Manufacturer | Model | Serial No | Date of manufacture/ Installation | Max voltage | Permanent filtration |
|---|---|---|---|---|---|---|
1 | X-ray Tube | |||||
Toshiba Electron Tubes and Devices Co., Ltd. | E7843X | 11G485 | N/A | 150 kVp | 1.3 A1/75 | |
Collimator | ||||||
GE HUALUN MEDICAL SYSTEMS Co. Ltd | 5189248 | 81622HL | December, 2011 | N/A | N/A | |
2 | X-ray Tube | |||||
Nanjing Perlove Medical Equipment Co. Ltd | HD052-30-50/125 | 1111132 | 2014 | 125kV | 2.5 mm Al | |
Collimator | ||||||
Nanjing Perlove Medical Equipment Co. Ltd | N/A | 1016 | N/A | 1.5 mm Al | ||
3 | X-ray Tube | |||||
IMD Generators s.r.l | E-100R HF XR03 C | 11/14-12346 | December, 2012 | 80 kVp | 1.8 mm Al | |
Collimator | ||||||
Ralco s.r.l | R105 | 1500471 | January, 2015 | N/A | 2 mm Al/75 | |
4 | X-ray Tube | |||||
SMAM arl | MX 4000Dual | 021036 | April, 2002 | 125 | 3.5 mm Al | |
Collimator | ||||||
Ralco | R221/A | 026/123 | March, 2008 | 2 mm Al | ||
5 | X-ray Tube | |||||
Nanjing Perlove Medical Equipment Co. Ltd | N/A | A1411138 | December, 2014 | 110 kV | 2.5 mm Al | |
Collimator | ||||||
Nanjing Perlove Medical Equipment Co. Ltd | X-101A | 1401199 | December, 2014 | N/A | 1.2 mm Al | |
Set mAs | Set kVp | Dose (µGy) | Convert µGy to mGy ( |
|
|---|---|---|---|---|
5 | 80 | 194.75 |
|
|
10 | 80 | 374.3 |
|
|
40 | 80 | 1311 |
|
|
60 | 80 | 1984.2 |
|
|
Set mAs | Set kVp | Dose (µGy) | Convert µGy to mGy ( |
|
|---|---|---|---|---|
5 | 80 | 267.0 |
|
|
10 | 80 | 569.4 |
|
|
20 | 80 | 1197.5 |
|
|
32 | 80 | 1932.5 |
|
|
40 | 80 | 2458.5 |
|
|
Set mAs | Set kVp | Dose (µGy) | Convert µGy to mGy ( |
|
|---|---|---|---|---|
5 | 80 | 244.2 |
|
|
10 | 80 | 460.2 |
|
|
20 | 80 | 933.7 |
|
|
32 | 80 | 1507.7 |
|
|
Set mAs | Set kVp | Dose (µGy) | Convert µGy to mGy ( |
|
|---|---|---|---|---|
5 | 80 | 321.93 |
|
|
10 | 80 | 626.6 |
|
|
20 | 80 | 1206 |
|
|
32 | 80 | 2094 |
|
|
Set mAs | Set kVp | Dose (µGy) | Convert µGy to mGy ( |
|
|---|---|---|---|---|
12 | 100 | 462.8 |
|
|
16 | 100 | 758 |
|
|
20 | 100 | 781.1 |
|
|
25 | 100 | 711.1 |
|
|
Set mAs | Set kVp | Max. kVp measured |
|
|---|---|---|---|
8 | 50 | 60.5 |
|
8 | 60 | 68.3 |
|
8 | 90 | 93.0 |
|
8 | 100 | 102.7 |
|
kVp accuracy = |
| ||
Set mAs | Set kVp | Max. kVp measured |
|
|---|---|---|---|
8 | 50 | 55.55 |
|
8 | 60 | 67.00 |
|
8 | 70 | 79.30 |
|
8 | 80 | 81.80 |
|
8 | 90 | 91.95 |
|
Average kVp error = |
| ||
Set mAs | Set kVp | Max. kVp measured |
|
|---|---|---|---|
8 | 60 | 63.23 |
|
8 | 70 | 75.94 |
|
8 | 80 | 87.38 |
|
8 | 90 | 88.37 |
|
8 | 100 | 98.37 |
|
Average kVp error = |
| ||
Set mAs | Set kVp | Max. kVp measured |
|
|---|---|---|---|
8 | 60 | 60.67 |
|
8 | 70 | 71.60 |
|
8 | 80 | 83.30 |
|
8 | 90 | 90.23 |
|
8 | 100 | 100.47 |
|
Average kVp error = |
| ||
Set mAs | Set kVp | Max. kVp measured |
|
|---|---|---|---|
12 | 70 | 69.8 |
|
12 | 80 | 79.5 |
|
12 | 90 | 89.7 |
|
12 | 100 | 99.8 |
|
kVp accuracy = |
| ||
Set mAs | Set kVp | Exposure measurements obtained |
|
| |
|---|---|---|---|---|---|
max kVp | Max. Mean kVp | ||||
8 | 70 | 77.0 | 77.0 | -7 | 49 |
8 | 70 | 76.0 | 76.1 | -6 | 36 |
8 | 70 | 77.0 | 77.1 | -7 | 49 |
8 | 70 | 77.8 | 77.9 | -7.8 | 60.84 |
|
| ||||
Set mAs | Set kVp | Exposure measurements obtained |
|
| |
|---|---|---|---|---|---|
max kVp | Max. Mean kVp | ||||
9 | 70 | 77.1 | 76.1 | -6.1 | 37.21 |
9 | 70 | 76.0 | 76.0 | -6.0 | 36.00 |
9 | 70 | 77.0 | 77.0 | -7.0 | 49.00 |
9 | 70 | 77.8 | 77.8 | -7.8 | 60.84 |
9 | 70 | 77.8 | 77.8 | -7.8 | 60.84 |
|
| ||||
Set mAs | Set kVp | Exposure measurements obtained |
|
| |
|---|---|---|---|---|---|
max kVp | Max. Mean kVp | ||||
8 | 70 | 79.3 | 79.3 | -9.3 | 86.49 |
8 | 70 | 74.7 | 74.7 | -4.7 | 22.09 |
8 | 70 | 75.0 | 75.0 | -5.0 | 25.00 |
8 | 70 | 75.5 | 75.5 | -5.5 | 30.25 |
8 | 70 | 75.2 | 75.2 | -5.2 | 27.04 |
|
| ||||
Set mAs | Set kVp | Exposure measurements obtained |
|
| |
|---|---|---|---|---|---|
max kVp | Max. Mean kVp | ||||
8 | 70 | 72.1 | 72.1 | -2.1 | 4.41 |
8 | 70 | 71.3 | 71.3 | -1.3 | 1.69 |
8 | 70 | 71.4 | 71.4 | -1.4 | 1.96 |
214.8 |
| ||||
Set mAs | Set kVp | Exposure measurements obtained |
|
| |
|---|---|---|---|---|---|
max kVp | Max. Mean kVp | ||||
8 | 70 | 71.4 | 70.7 | -0.7 | 0.49 |
8 | 70 | 71.4 | 70.7 | -0.7 | 0.49 |
8 | 70 | 71.4 | 69.1 | 0.9 | 0.81 |
8 | 70 | 71.2 | 69.8 | 0.2 | 0.04 |
8 | 70 | 71.3 | 69.1 | 0.9 | 0.81 |
|
| ||||
QC Parameters | Acceptable limits | 1 | 2 | 3 | 4 | 5 |
|---|---|---|---|---|---|---|
mAs Linearity | ≤ ± 0.1 or 10% | 0.086 | 0.071 | 0.032 | 0.041 | 0.157 |
kVp Accuracy | ≤ ± 5% | 10 | 8 | 5 | 1.7 | 0.2 |
Reproducibility | ≤ ± 5% | 10 | 10 | 9 | 2.8 | 1.2 |
PRD | Personal Radiation Detector |
mAs | Milliampere seconds |
kVp | Kilovoltage |
CoL | Coefficient of Linearity |
mR | Millroentgen |
SD () | Standard Deviation |
CV | Coefficient of Variation |
µGy | Microgray |
mGy | Milligray |
| [1] | A. Barzylovych, Y. Ursakii, A. Nadezhdenko, T. Mamatova, O. Chykarenko, and S. Kravchenko, "The influence of medical services public management on the population’life quality," 2021. |
| [2] | A. Mohammad Mosadeghrad, "Healthcare service quality: towards a broad definition," International journal of health care quality assurance, vol. 26, pp. 203-219, 2013. |
| [3] | S. Code, "Safety Procedures for the Installation, Use and Control of X-ray Equipment in Large Medical Radiological Facilities," 2024. |
| [4] | M. Amurao, D. A. Gress, M. A. Keenan, P. H. Halvorsen, J. A. Nye, and M. Mahesh, "Quality management, quality assurance, and quality control in medical physics," Journal of applied clinical medical physics, vol. 24, p. e13885, 2023. |
| [5] | H. Delis, K. Christaki, B. Healy, G. Loreti, G. L. Poli, P. Toroi, et al., "Moving beyond quality control in diagnostic radiology and the role of the clinically qualified medical physicist," Physica Medica, vol. 41, pp. 104-108, 2017. |
| [6] | G. K. Korir, J. S. Wambani, and I. K. Korir, "Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology," SA Journal of Radiology, vol. 15, pp. 70-79, 2011. |
| [7] | L. L. Nkuba and P. B. Nyanda, "Compliance and Quality Control Monitoring of Diagnostic X ray Facilities in Dar es Salaam City, Tanzania," Brazilian Journal of Radiation Sciences, vol. 5, 2017. |
| [8] | M. A. Ali, S. A. Vuai, S. A. Suleiman, and M. F. Mwema, "Assessment of regulatory compliance in diagnostic X-ray facilities based on a radiation safety audit in Zanzibar," Journal of Medical Imaging and Radiation Sciences, vol. 57, p. 102190, 2026. |
| [9] | J. Alyami and M. Nassef, "Assessment of diagnostic radiology facilities technical radiation protection requirements in KSA," Applied Sciences, vol. 12, p. 7284, 2022. |
| [10] | B. Van der Merwe, M. Nel, and S. Kruger, "Radiation safety requirements for training of users of diagnostic X-ray equipment in South Africa," African Journal of Health Professions Education, vol. 9, pp. 123-127, 2017. |
| [11] | A. Sonawane, M. Singh, J. S. Kumar, A. Kulkarni, V. Shirva, and A. Pradhan, "Radiological safety status and quality assurance audit of medical X-ray diagnostic installations in India," Journal of Medical Physics, vol. 35, pp. 229-234, 2010. |
| [12] | M. Sandborg, J. N. Althén, and A. Gustafsson, "Efficient quality assurance programs in radiology and nuclear medicine in Östergötland, Sweden," Radiation protection dosimetry, vol. 139, pp. 410-417, 2010. |
| [13] | V. Patel, G. Sindhwani, M. Gupta, S. Arora, A. Mishra, J. Bhatt, et al., "A comprehensive approach towards quality and safety in diagnostic imaging services: our experience at a rural tertiary health care center," Journal of Clinical and Diagnostic Research: JCDR, vol. 11, p. TC10, 2017. |
| [14] | B. Allen, M. Chatfield, J. Burleson, and W. T. Thorwarth, "Improving diagnosis in health care: perspectives from the American College of Radiology," Diagnosis, vol. 4, pp. 113-124, 2017. |
| [15] | M. Lötter, "Development and evaluation of a quality assurance programme for a radiology department," 2018. |
| [16] | J. C. Broder, S. F. Cameron, W. T. Korn, and S. J. Baccei, "Creating a radiology quality and safety program: principles and pitfalls," Radiographics, vol. 38, pp. 1786-1798, 2018. |
| [17] | I. P. 154, "Optimisation of radiological protection in digital radiology techniques for medical imaging," Annals of the ICRP, vol. 52, pp. 11-145, 2023. |
| [18] | C. Cloete, "An activity-based workload modelling approach for determining diagnostic radiographer staffing requirements within a diagnostic radiology practice," Stellenbosch: Stellenbosch University, 2021. |
| [19] | D. Polevoy, D. Kazimirov, M. Gilmanov, and D. Nikolaev, "No reproducibility, No progress: rethinking CT benchmarking," Journal of Imaging, vol. 11, p. 344, 2025. |
| [20] | T. Chusin, R. Wongchai, S. Moonkham, T. Pengpan, and K. Aphiwatthanasumet, "Evaluation of the reproducibility of automatic exposure control systems in general X-ray machines using a coin-based method: " Radiological Physics and Technology, vol. 18, pp. 1239-1246, 2025. |
| [21] | N. I. Barau, W. E. Mangset, I.-O. M. Ijeoma, and Y. B. Mohammed, "Evaluation of Quality Control Standards in Medical Diagnostic X-Ray Facilities in North-Eastern Nigeria," Nigerian Journal of Physics, vol. 34, pp. 110-117, 2025. |
| [22] | A. Pilipović, G. Baršić, M. Katić, and M. Rujnić Havstad, "Repeatability and reproducibility assessment of a PolyJet technology using x-ray computed tomography," Applied Sciences, vol. 10, p. 7040, 2020. |
| [23] | D. Prakash and R. P. Kotian, "Standard X-Ray Tubes: Basic Principles, Types of X-Ray Tubes, and Routine Quality Control," in Fundamentals of X-ray Imaging: Basic Principles, Quality Control, Clinical Applications, and Safety, ed: Springer, 2025, pp. 87-184. |
| [24] | M. Anburajan and J. K. Sharma, "“Overview of X-Ray Tube Technology” 21," Biomedical engineering and its applications in healthcare, p. 519, 2019. |
| [25] | D. Meikle and G. Hill, "X‐ray Production," Practical Radiotherapy: Physics and Equipment, 3rd Edition, pp. 23-33, 2019. |
| [26] | R. Pauwels, O. Silkosessak, R. Jacobs, R. Bogaerts, H. Bosmans, and S. Panmekiate, "A pragmatic approach to determine the optimal kVp in cone beam CT: balancing contrast-to-noise ratio and radiation dose," Dentomaxillofacial Radiology, vol. 43, p. 20140059, 2014. |
| [27] | E. Seeram, R. Davidson, S. Bushong, and H. Swan, "Radiation dose optimization research: Exposure technique approaches in CR imaging-A literature review," Radiography, vol. 19, pp. 331-338, 2013. |
| [28] | G. X. Ding and P. Munro, "Radiation exposure to patients from image guidance procedures and techniques to reduce the imaging dose," Radiotherapy and Oncology, vol. 108, pp. 91-98, 2013. |
| [29] | C. R. Kim and J. Y. Jeon, "Radiation dose and image conspicuity comparison between conventional 120 kVp and 150 kVp with spectral beam shaping for temporal bone CT," European Journal of Radiology, vol. 102, pp. 68-73, 2018. |
| [30] | A. A. Oglat, "Comparison of X-ray films in term of kVp, mA, exposure time and distance using Radiographic Chest Phantom as a radiation quality," Journal of Radiation Research and Applied Sciences, vol. 15, p. 100479, 2022. |
| [31] | T. L. Fauber, T. F. Cohen, and M. C. Dempsey, "High kilovoltage digital exposure techniques and patient dosimetry," Radiologic technology, vol. 82, 2011. |
| [32] | A. Almutairi and A. A. Mohammed, "Performance assessment of GE conventional X-ray System installed in Radiological Science Department of King Saud University," Am. J. PharmTcch. Res, vol. 14, 2024. |
| [33] | M. Hashemi, S. Bayani Roodi, F. Shahedi, M. Momennezhad, H. Zare, and H. Gholamhosseinian, "Quality assessment of conventional X-ray diagnostic equipment by measuring X-ray exposure and tube output parameters in Great Khorasan Province, Iran," Iranian Journal of Medical Physics, vol. 16, pp. 34-40, 2019. |
| [34] | R. A. Lynch, T. Smith, M. C. Jacobs, S. J. Frank, and K. J. Kearfott, "A radiation weather station: development of a continuous monitoring system for the collection, analysis, and display of environmental radiation data," Health Physics, vol. 115, pp. 590-599, 2018. |
| [35] | S. Pommé, J. Paepen, and R. Van Ammel, "Linearity check of an ionisation chamber through 99 mTc half-life measurements," Applied Radiation and Isotopes, vol. 140, pp. 171-178, 2018. |
| [36] | J. P. Bissonnette, D. J. Moseley, and D. A. Jaffray, "A quality assurance program for image quality of cone‐beam CT guidance in radiation therapy," Medical physics, vol. 35, pp. 1807-1815, 2008. |
| [37] | F. Zarb, L. Rainford, and M. F. McEntee, "Image quality assessment tools for optimization of CT images," Radiography, vol. 16, pp. 147-153, 2010. |
| [38] | H. Herath, H. Herath, N. Madusanka, and B.-I. Lee, "A systematic review of medical image quality assessment," Journal of Imaging, vol. 11, p. 100, 2025. |
| [39] | B. Ijabor, C. Nzotta, and A. Omojola, "Quality control test of conventional X-Ray systems in Delta State, South-South, Nigeria," World Scientific News, vol. 157, pp. 140-153, 2021. |
| [40] | M. P. Kadima, "Assessment of quality control in medical diagnostic x-ray facilities in the western region of Kenya," 2013. |
| [41] | U. Hirvonen, "X-Ray Compliance Testing of a Mammography System on Manufacturing Line: Dose Reproducibility & Linearity," 2015. |
| [42] | P. T. Hornsby, "Simplified Quality Control System in Contrast to Performance Measurements to Assure Minimized Patient Exposure From Diagnostic X-Ray Machines," Idaho State University, 2024. |
| [43] | I. I. Suliman, "Patient dosimetry and quality control in diagnostic radiology," Sudan Academy of Sciences, Khartoum (Sudan)2007. |
| [44] | E. Karavas, B. Ece, S. Aydın, M. Kocak, Z. Cosgun, I. E. Bostanci, et al., "Are we aware of radiation: A study about necessity of diagnostic X-ray exposure," World Journal of Methodology, vol. 12, p. 264, 2022. |
| [45] | E. A. Alakrout, M. M. Zayed, A. A. Al-Lisini, and H. A. Al-Tawerghi, "Public Awareness and Knowledge of Ionizing Radiation Risks from Medical Imaging Examinations in Western Libya," Surman Journal of Science and Technology, vol. 7, pp. 105-117, 2025. |
| [46] | J. Ford, "Radiation, people and the environment," 2004. |
| [47] | S. Visweswaran, K. Kanagaraj, S. Joseph, and V. Perumal, "Medical imaging: Contribution toward background radiation and human exposure," Journal of Radiation and Cancer Research, vol. 9, pp. 177-182, 2018. |
| [48] | A. R. A. Mishal, M. S. Jabr, and Z. I. Muhammad, "Radiation: Types, Natural sources, and Industrial Sources," Current Clinical and Medical Education, vol. 2, pp. 89-94, 2024. |
| [49] | V. Harahap, J. Saragih, and E. B. Ginting, "IMPLEMENTATION OF RADIATION PROTECTION QUALITY CONTROL IN RADIOLOGY INSTALLATIONS IN ACCORDANCE WITH MINISTRY OF HEALTH REGULATION NO. 8 OF 2011," Jurnal Ilmiah METADATA, vol. 5, pp. 456-478, 2023. |
| [50] | G. Fan and Q. Wang, "Quality control and quality assurance," in Clinical molecular diagnostics, ed: Springer, 2021, pp. 97-113. |
| [51] | H. H. van Rossum, "Technical quality assurance and quality control for medical laboratories: a review and proposal of a new concept to obtain integrated and validated QA/QC plans," Critical reviews in clinical laboratory sciences, vol. 59, pp. 586-600, 2022. |
APA Style
Kamara, S. (2026). Evaluation of mAs Linearity, kVp Accuracy, and Reproducibility in Selected X-ray Diagnostic Facilities in Freetown. Radiation Science and Technology, 12(2), 31-43. https://doi.org/10.11648/j.rst.20261202.12
ACS Style
Kamara, S. Evaluation of mAs Linearity, kVp Accuracy, and Reproducibility in Selected X-ray Diagnostic Facilities in Freetown. Radiat. Sci. Technol. 2026, 12(2), 31-43. doi: 10.11648/j.rst.20261202.12
@article{10.11648/j.rst.20261202.12,
author = {Saidu Kamara},
title = {Evaluation of mAs Linearity, kVp Accuracy, and Reproducibility in Selected X-ray Diagnostic Facilities in Freetown},
journal = {Radiation Science and Technology},
volume = {12},
number = {2},
pages = {31-43},
doi = {10.11648/j.rst.20261202.12},
url = {https://doi.org/10.11648/j.rst.20261202.12},
eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.rst.20261202.12},
abstract = {Quality control in medical radiology is a process used by a regulatory authority to evaluate the performance of a diagnostic X-ray machine, including acceptable limits and the actions necessary to meet regulatory requirements. This study evaluates the radiation dose outputs of X-ray equipment in diagnostic medical hospitals by measuring mAs linearity, kVp accuracy, and reproducibility as quality control parameters to ensure that radiation doses administered to patients meet standard regulatory requirements. Radiation measuring instruments were used in this work to measure mAs, kVp accuracy, and reproducibility outputs for five (5) medical diagnostic X-ray machines. The detector was placed on the couch at a distance of 100 cm from the collimator to take measurements for each parameter under investigation. Dose rate measurements for each parameter were recorded one at a time. The mathematical equations for mAs linearity, kVp accuracy, and reproducibility were used to compute the linearity coefficient, percentage error, and coefficient of variation. The computed figures for mAs linearity for the five X-ray machines show that four (4) passed and one fell below the allowable limit. Data for kVp accuracy indicate that three of the machines are within the allowable limits and two (2) failed, while the calculated values for reproducibility show that only two (2) machines passed the tests. The study generates empirical quality control data rarely reported in X-ray diagnostic centers. It verifies the linearity coefficient, percentage kVp error, and the coefficient of variation with practical knowledge in radiology studies. Additionally, it provides a systematic mathematical approach to compute mAs linearity, kVp accuracy, and reproducibility for teaching radiology in non-diagnostic laboratory settings.},
year = {2026}
}
TY - JOUR T1 - Evaluation of mAs Linearity, kVp Accuracy, and Reproducibility in Selected X-ray Diagnostic Facilities in Freetown AU - Saidu Kamara Y1 - 2026/10/21 PY - 2026 N1 - https://doi.org/10.11648/j.rst.20261202.12 DO - 10.11648/j.rst.20261202.12 T2 - Radiation Science and Technology JF - Radiation Science and Technology JO - Radiation Science and Technology SP - 31 EP - 43 PB - Science Publishing Group SN - 2575-5943 UR - https://doi.org/10.11648/j.rst.20261202.12 AB - Quality control in medical radiology is a process used by a regulatory authority to evaluate the performance of a diagnostic X-ray machine, including acceptable limits and the actions necessary to meet regulatory requirements. This study evaluates the radiation dose outputs of X-ray equipment in diagnostic medical hospitals by measuring mAs linearity, kVp accuracy, and reproducibility as quality control parameters to ensure that radiation doses administered to patients meet standard regulatory requirements. Radiation measuring instruments were used in this work to measure mAs, kVp accuracy, and reproducibility outputs for five (5) medical diagnostic X-ray machines. The detector was placed on the couch at a distance of 100 cm from the collimator to take measurements for each parameter under investigation. Dose rate measurements for each parameter were recorded one at a time. The mathematical equations for mAs linearity, kVp accuracy, and reproducibility were used to compute the linearity coefficient, percentage error, and coefficient of variation. The computed figures for mAs linearity for the five X-ray machines show that four (4) passed and one fell below the allowable limit. Data for kVp accuracy indicate that three of the machines are within the allowable limits and two (2) failed, while the calculated values for reproducibility show that only two (2) machines passed the tests. The study generates empirical quality control data rarely reported in X-ray diagnostic centers. It verifies the linearity coefficient, percentage kVp error, and the coefficient of variation with practical knowledge in radiology studies. Additionally, it provides a systematic mathematical approach to compute mAs linearity, kVp accuracy, and reproducibility for teaching radiology in non-diagnostic laboratory settings. VL - 12 IS - 2 ER -