Research Article | | Peer-Reviewed

Evaluation of mAs Linearity, kVp Accuracy, and Reproducibility in Selected X-ray Diagnostic Facilities in Freetown

Received: 13 September 2026     Accepted: 27 September 2026     Published: 21 October 2026
Views:       Downloads:
Abstract

Quality control in medical radiology is a process used by a regulatory authority to evaluate the performance of a diagnostic X-ray machine, including acceptable limits and the actions necessary to meet regulatory requirements. This study evaluates the radiation dose outputs of X-ray equipment in diagnostic medical hospitals by measuring mAs linearity, kVp accuracy, and reproducibility as quality control parameters to ensure that radiation doses administered to patients meet standard regulatory requirements. Radiation measuring instruments were used in this work to measure mAs, kVp accuracy, and reproducibility outputs for five (5) medical diagnostic X-ray machines. The detector was placed on the couch at a distance of 100 cm from the collimator to take measurements for each parameter under investigation. Dose rate measurements for each parameter were recorded one at a time. The mathematical equations for mAs linearity, kVp accuracy, and reproducibility were used to compute the linearity coefficient, percentage error, and coefficient of variation. The computed figures for mAs linearity for the five X-ray machines show that four (4) passed and one fell below the allowable limit. Data for kVp accuracy indicate that three of the machines are within the allowable limits and two (2) failed, while the calculated values for reproducibility show that only two (2) machines passed the tests. The study generates empirical quality control data rarely reported in X-ray diagnostic centers. It verifies the linearity coefficient, percentage kVp error, and the coefficient of variation with practical knowledge in radiology studies. Additionally, it provides a systematic mathematical approach to compute mAs linearity, kVp accuracy, and reproducibility for teaching radiology in non-diagnostic laboratory settings.

Published in Radiation Science and Technology (Volume 12, Issue 2)
DOI 10.11648/j.rst.20261202.12
Page(s) 31-43
Creative Commons

This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited.

Copyright

Copyright © The Author(s), 2026. Published by Science Publishing Group

Keywords

kVp Accuracy, mAs Linearity, Quality Control, Radiology, Reproducibility, X-ray

1. Introduction
The global implementation of health services has improved rapidly, but much more remains to be addressed, especially the quality of services due to decentralization. Public health and the quality of life of people will be improved by increasing the quality and efficiency of health services . The quality assurance policy in healthcare services serves as a guide for implementing quality healthcare services . Radiology is a health service meant to support the quality of health. The ability and quality of service vary in both government and private radiological health facilities and have not fully met the standard regulatory requirements. Quality control tests are normally conducted so that radiological facilities meet the regulatory requirements . A compliance test in a diagnostic health facility is carried out to make sure that X-ray machines meet radiation safety requirements and provide accurate information . Precise and accurate patient diagnosis is the main objective for any quality assurance program at the installation of a radiology facility and comprehensively involves reducing radiation exposure, improving diagnostic images, and reducing costs.
Quality in diagnostic radiology is when the expected objectives or outputs are met and always align with new developments in professional knowledge . The inputs, process, and output components must be highlighted/detailed in the quality assurance program. One of the quality assurance activities is quality control. Studies have reported the significance of quality control tests in diagnostic radiology to ensure accurate exposure parameters, production of quality images, and radiation protection of patients in line with acceptable international standards and the regulations of national regulatory bodies. The problems of quality control of diagnostic X-ray machines are common in developing countries such as Tanzania, Nigeria, Sierra Leone, etc according to various research evidence The main objective of this research is to evaluate the reproducibility, kVp accuracy, and mAs linearity parameters in some X-ray diagnostic facilities in the capital city, Freetown, Sierra Leone. Reproducibility highlights the X-ray machine's consistency. It is used to check the consistency of the radiation output . This is done by keeping the mA and time fixed and measuring the radiation output at various kV settings. kVp accuracy is the accelerating voltage. kVp stands for kilovoltage peak, and it is the maximum voltage difference between the cathode and the anode in an X-ray tube . The electrons are accelerated by this voltage from the cathode to the anode, which controls the energy (quality) and the penetration of the resulting X-ray beam; no X-rays would be produced in the absence of this voltage. An increase in kVp increases the average energy of the photons. The number of photons produced is also determined by the kVp. The higher the kVp, the higher the number of photons produced. Increasing kVp increases the beam's quality and quantity, as well as the patient dose and receptor exposure . The quality of the X-ray beam and exposure to the patient are affected by the peak potential of the X-ray generator. The tube potential is assessed to make sure that the resulting kVp is very close to the one set on the equipment by the operator of the X-ray equipment. A linearity test is conducted to verify the linearity of the radiation output with respect to changes in timer stations, while maintaining the mA station constant at a particular kV station . The purpose of performing linearity tests is to monitor the patient dose and image quality . Variation in output intensity during diagnostic exposures can result in unnecessary dose to the patient due to repeats from poor-quality images . While performing this test, the assumption is that every other factor, such as kVp and timer, is working accurately. It is important to perform QC tests on the kVp as well as the timer before completing the linearity test, as if there are variations in linearity, mAs may not be the culprit . Inaccurate performance from the kVp meter and X-ray timer can also produce the same results. Keeping exposure time and kVp constant, radiation output is measured at different mA stations. Measurement for the mA loading station is to be repeated a number of times each to eliminate statistical errors. Radiation output readings of each mA loading station are averaged, and the coefficient of linearity (CoL) is evaluated from average mR/mAs or mGy/mAs .
The use of radiation in medical applications has rapidly increased over the years. One major public exposure to radiation is diagnostic X-ray examinations, and it is the largest man-made source of ionizing radiation in the world. The most commonly used X-rays for medical imaging are X-ray radiography, computed tomography (CT), mammography, angiography, and fluoroscopy. Medical exposures can best be justified and optimized by adequately implementing Quality Control (QC) programs . In Sierra Leone, Diagnostic X-ray radiography is the most commonly used ionizing radiation in medical facilities and has played a key role in diagnosing diseases, injuries, and in oncology (radiation therapy). The availability and utilization of X-ray equipment in both government and private health facilities are increasing globally. To prevent high doses, radiation leakage, and the damaging effects of radiation on the public and occupationally exposed workers (radiation workers), it is essential to maintain adequate quality control (QC) standards in X-ray examinations.
2. Materials and Methods
In general, the materials used in the process included a survey meter, personal radiation detectors (PRD), quality control kits (tape rule, PTW Diavolt kVp and dose meter, and cone), control console, and a collimator. Radiographic tests/measurements of three major quality control parameters (milliampere-seconds (mAs) linearity, kilovoltage peak (kVp) accuracy, and reproducibility) were carried out on five (5) X-ray machines in different diagnostic hospitals in Freetown. Measurements of the three sets of parameters were performed by placing the detector (kVp diavolt dose meter) on the couch at a distance of 100 cm (1m) from the collimator (source). Before making the exposures, the collimator shutters were fully closed, and the X-ray tube was pointed downward. The mAs and kVp values for the three different parameters (reproducibility, kVp accuracy, and mAs linearity) were selected for all X-ray equipment at the different medical facilities. The selected mAs and kVp reproducibility values are unique for four (4) of the X-ray machines, with one X-ray machine showing a variation in its selected mAs value relative to the others. X-ray kVp reproducibility tests require keeping kVp and mAs exposure parameters fixed across consecutive measurements to assess output stability. For kVp-accuracy dose measurements, the set mAs was kept constant across various kVp values. A similar procedure was done for mAs linearity, where kVp was kept constant at different mAs. The dose measurements for the different parameters were recorded and tabulated for each X-ray equipment. The data obtained for each parameter were then computed and analyzed.
2.1. Mathematical Equations
Below are the three quality control parameters and their mathematical expressions used to compute the collected data:
2.1.1. mAs Linearity
The coefficient of linearity (CoL) is evaluated from average mR/mAs or mGy/mAs as follows:
mAs Linearity = maxDose - minDosemaxDose +minDose
Tolerance: Coefficient of timer linearity ≤ ± 0.1 or 10%
2.1.2. kVp Accuracy
Percentage kVp error = kVpmeasured- kVpsetkVpset ×100%
The allowed standard kVp limit = ± 5
2.2.3. Reproducibility
The average (X) of mR/mAs or mGy/mAs is calculated. The consistency at each kV setting is determined by calculating the coefficient of variation.
Coefficient of Variation (CV) = Standard Deviation σMean x̅ ×100%
Standard Deviation σ = 1n-1∑i-1nxi -x12
The standard reproducibility limit for a diagnostic X-ray machine = ± 5%
The table below provides key information to describe the characteristics of the five X-ray machines:
Table 1. Characteristics of X-ray machines.

No

Manufacturer

Model

Serial No

Date of manufacture/ Installation

Max voltage

Permanent filtration

1

X-ray Tube

Toshiba Electron Tubes and Devices Co., Ltd.

E7843X

11G485

N/A

150 kVp

1.3 A1/75

Collimator

GE HUALUN MEDICAL SYSTEMS Co. Ltd

5189248

81622HL

December, 2011

N/A

N/A

2

X-ray Tube

Nanjing Perlove Medical Equipment Co. Ltd

HD052-30-50/125

1111132

2014

125kV

2.5 mm Al

Collimator

Nanjing Perlove Medical Equipment Co. Ltd

N/A

1016

N/A

1.5 mm Al

3

X-ray Tube

IMD Generators s.r.l

E-100R HF XR03 C

11/14-12346

December, 2012

80 kVp

1.8 mm Al

Collimator

Ralco s.r.l

R105

1500471

January, 2015

N/A

2 mm Al/75

4

X-ray Tube

SMAM arl

MX 4000Dual

021036

April, 2002

125

3.5 mm Al

Collimator

Ralco

R221/A

026/123

March, 2008

2 mm Al

5

X-ray Tube

Nanjing Perlove Medical Equipment Co. Ltd

N/A

A1411138

December, 2014

110 kV

2.5 mm Al

Collimator

Nanjing Perlove Medical Equipment Co. Ltd

X-101A

1401199

December, 2014

N/A

1.2 mm Al

3. Results and Discussions
3.1. Results
3.1.1. mAs Linearity Tests
Table 2. mAs linearity measurements for X-ray 1.

Set mAs

Set kVp

Dose (µGy)

Convert µGy to mGy (μGy1000)

mGymAs

5

80

194.75

194∙751000=0∙19475

0.194755=0∙03895

10

80

374.3

374∙31000=0∙3743

0.374310=0∙03743

40

80

1311

13111000=1∙311

1∙31140=0∙032775

60

80

1984.2

1984∙21000=1∙9842

1∙984260=0∙03307

mAs Linearity = maxDose - minDosemaxDose +minDose= 0.03895 - 0.0327750.03895 +0.032775= 0.08609≈0.086
Table 3. mAs linearity measurements for X-ray 2.

Set mAs

Set kVp

Dose (µGy)

Convert µGy to mGy (μGy1000)

mGymAs

5

80

267.0

267∙01000=0∙267

0.2675=0∙0534

10

80

569.4

569∙91000=0∙5699

0.569910=0∙0569

20

80

1197.5

1197∙51000=1∙1975

1∙197520=0∙0599

32

80

1932.5

1932∙51000=1∙932

1∙93232=0∙0604

40

80

2458.5

2458∙510002∙459

2∙45940=0∙0615

mAs Linearity = maxDose - minDosemaxDose +minDose= 0.0615 - 0.05340.0615 +0.0534=0.00810.1149= 0.07049≈0.071
Table 4. mAs linearity measurements for X-ray 3.

Set mAs

Set kVp

Dose (µGy)

Convert µGy to mGy (μGy1000)

mGymAs

5

80

244.2

245∙21000=0∙2452

0.24525=0∙04904

10

80

460.2

460∙21000=0∙46017

0.4601710=0∙04602

20

80

933.7

933∙71000=0∙9337

0∙933720=0∙04669

32

80

1507.7

1507∙71000=1∙5077

1∙507732=0∙04712

mAs Linearity = maxDose - minDosemaxDose +minDose= 0.0490 - 0.04600.0490 +0.0460=0.00300.0950= 0.03158≈0.032
Table 5. mAs linearity measurements for X-ray 4.

Set mAs

Set kVp

Dose (µGy)

Convert µGy to mGy (μGy1000)

mGymAs

5

80

321.93

321∙931000=0∙32193

0.321935=0∙0644

10

80

626.6

626∙61000=0∙6266

0.626610=0∙0627

20

80

1206

12061000=1∙206

1∙20620=0∙0603

32

80

2094

20941000=2∙094

2∙09432=0∙0654

mAs Linearity = maxDose - minDosemaxDose +minDose= 0.0654 - 0.06030.0654 +0.0603=0.00510.1257= 0.04057≈0.041
Table 6. mAs linearity measurements for X-ray 5.

Set mAs

Set kVp

Dose (µGy)

Convert µGy to mGy (μGy1000)

mGymAs

12

100

462.8

462∙81000=0∙4628

0.462812=0∙038557

16

100

758

7581000=0∙578

0.57816=0∙036125

20

100

781.1

781∙11000=0∙7811

0∙781120=0∙039055

25

100

711.1

711∙11000=0∙7111

0∙711125=0∙02844

mAs Linearity = maxDose - minDosemaxDose +minDose= 0.039055 - 0.028440.039055 +0.02844=0.01060.0675= 0.15727≈0.157
3.1.2. kVp accuracy Tests
Table 7. kVp accuracy measurements for X-ray 1.

Set mAs

Set kVp

Max. kVp measured

% kVp error = kVpmeasured- kVpsetkVpset ×100%

8

50

60.5

%kVp error = 60.5-5050×100% =21∙00%

8

60

68.3

%kVp error = 68.3-6060×100% =13∙00%

8

90

93.0

%kVp error = 60.5-5050×100% =3∙33%

8

100

102.7

%kVp error = 60.5-5050×100% =2∙70%

kVp accuracy =

21.00+13.83+3.33+2.704=10.21≈10%

Table 8. kVp accuracy measurements for X-ray 2.

Set mAs

Set kVp

Max. kVp measured

% kVp error = kVpmeasured- kVpsetkVpset ×100%

8

50

55.55

%kVp error = 55.55-5050×100% =11∙10%

8

60

67.00

%kVp error = 67.0-6060×100% =11∙60%

8

70

79.30

%kVp error = 79.3-7070×100% =13∙28%

8

80

81.80

%kVp error = 81.8-8080×100% =2∙25%

8

90

91.95

%kVp error = 91.95-9090×100% =2∙17%

Average kVp error =

11.10+11.60+13.28+2.25+2.175=8.08≈8%

Table 9. kVp accuracy measurements for X-ray 3.

Set mAs

Set kVp

Max. kVp measured

% kVp error = kVpmeasured- kVpsetkVpset ×100%

8

60

63.23

%kVp error = 63.23-6060×100% =5∙38%

8

70

75.94

%kVp error = 75.94-7070×100% =8∙49%

8

80

87.38

%kVp error = 87.38-8080×100% =9∙22%

8

90

88.37

%kVp error = 88.37-9090×100% =1∙81%

8

100

98.37

%kVp error = 98.37-100100×100% =1∙63%

Average kVp error =

5.38+8.49+9.22+1.81+1.635=5.306≈5%

Table 10. kVp accuracy measurements for X-ray 4.

Set mAs

Set kVp

Max. kVp measured

% kVp error = kVpmeasured- kVpsetkVpset ×100%

8

60

60.67

%kVp error = 60.67-6060×100% =1∙12%

8

70

71.60

%kVp error = 71.6-7070×100% =2∙29%

8

80

83.30

%kVp error = 83.3-8080×100% =4∙13%

8

90

90.23

%kVp error = 90.23-9090×100% =0∙65%

8

100

100.47

%kVp error = 100.47-100100×100% =0∙47%

Average kVp error =

1.12+2.29+4.13+0.65+0.475=1.65≈1.7%

Table 11. kVp accuracy measurements for X-ray 5.

Set mAs

Set kVp

Max. kVp measured

% kVp error = kVpmeasured- kVpsetkVpset ×100%

12

70

69.8

%kVp error = 69.8-7070×100% =-0∙28%

12

80

79.5

%kVp error = 79.5-8080×100% =-0∙01%

12

90

89.7

%kVp error = 89.7-9090×100% =-0∙33%

12

100

99.8

%kVp error = 99.8-100100×100% =-0∙20%

kVp accuracy =

(-0.28)+(-0.01)+(-0.33)+(-0.20)4=-0.21≈-0.2%

3.1.3. Reproducibility Tests
Table 12. Reproducibility measurements for X-ray 1.

Set mAs

Set kVp x

Exposure measurements obtained

x - x1

x - x12

max kVp

Max. Mean kVp x1

8

70

77.0

77.0

-7

49

8

70

76.0

76.1

-6

36

8

70

77.0

77.1

-7

49

8

70

77.8

77.9

-7.8

60.84

∑x1=307.8 

∑xi-x12 =194.84

Standard Deviation σ = 1n-1∑i-1nxi -x12 = 14-1(194.84)=8.05
Mean x̅=∑x1n=307.84=76.95
Coefficient of Variation = S D σMean x̅ ×100%= 8.0576.95 ×100%=10%
Table 13. Reproducibility measurements for X-ray 2.

Set mAs

Set kVp x

Exposure measurements obtained

x - x1

x - x12

max kVp

Max. Mean kVp x1

9

70

77.1

76.1

-6.1

37.21

9

70

76.0

76.0

-6.0

36.00

9

70

77.0

77.0

-7.0

49.00

9

70

77.8

77.8

-7.8

60.84

9

70

77.8

77.8

-7.8

60.84

∑x1= 384.73

∑x-x12 =243.89

Standard Deviation σ = 1n-1∑i-1nxi -x12= 15-1(194.84)=7.80
Mean x̅=∑x1n=384.735=76.95
C V = S D σMean x̅ ×100%= 7.8076.95 ×100%=0.10 ×100%=10%
Table 14. Reproducibility measurements for X-ray 3.

Set mAs

Set kVp x

Exposure measurements obtained

x - x1

x - x12

max kVp

Max. Mean kVp x1

8

70

79.3

79.3

-9.3

86.49

8

70

74.7

74.7

-4.7

22.09

8

70

75.0

75.0

-5.0

25.00

8

70

75.5

75.5

-5.5

30.25

8

70

75.2

75.2

-5.2

27.04

∑x̅1=379.7 

∑x-x12 =190.87

Standard Deviation σ = 1n-1∑i-1nxi -x12= 15-1(190.87)=6.91
Mean x̅=∑x1n=379.75=75.94
C V = S D σMean x̅ ×100%= 6.9175.94 ×100%=0.090 ×100%=9%
Table 15. Reproducibility measurements for X-ray 4.

Set mAs

Set kVp x

Exposure measurements obtained

x - x1

x - x12

max kVp

Max. Mean kVp x1

8

70

72.1

72.1

-2.1

4.41

8

70

71.3

71.3

-1.3

1.69

8

70

71.4

71.4

-1.4

1.96

∑x1= 214.8

∑xi-x12 =8.06

Standard Deviation σ = 1n-1∑i-1nxi -x12= 13-1(8.06)=2.01
Mean x̅=∑x1n=214.83=71.6
C V = S D σMean x̅ ×100%= 2.0171.6 ×100%=0.028 ×100%=2.8%
Table 16. Reproducibility measurements for X-ray 5.

Set mAs

Set kVp x

Exposure measurements obtained

x - x1

x - x12

max kVp

Max. Mean kVp x1

8

70

71.4

70.7

-0.7

0.49

8

70

71.4

70.7

-0.7

0.49

8

70

71.4

69.1

0.9

0.81

8

70

71.2

69.8

0.2

0.04

8

70

71.3

69.1

0.9

0.81

∑x1=349.4 

∑xi-x12 =2.64

Standard Deviation σ = 1n-1∑i-1nxi -x12= 15-1(2.64)=0.81
Mean x̅=∑x1n=349.45=69.88
C V = S D σMean x̅ ×100%= 0.8169.88 ×100%=0.012 ×100%=1.2%
Summary data table
Table 17. Summary data table.

QC Parameters

Acceptable limits

1

2

3

4

5

mAs Linearity

≤ ± 0.1 or 10%

0.086

0.071

0.032

0.041

0.157

kVp Accuracy

≤ ± 5%

10

8

5

1.7

0.2

Reproducibility

≤ ± 5%

10

10

9

2.8

1.2

3.2. Discussion
Three quality control tests were carried out on five (5) X-ray machines in selected diagnostic medical hospitals in Freetown. Before the tests, a radiation survey was carried out around the X-ray installation at various locations, such as the door, console, patient waiting area, and lead barrier, to ensure the safe protection of workers, the general public, and the environment. The calculated figures obtained for mAs linearity are 0.086, 0.071, 0.032, 0.041, and 0.2. The allowable limit is ≤ ± 0.1 or 10%, which indicates that four (4) X-ray machines passed and one (1) failed the mAs linearity test. The kVp accuracy results are 10%, 8%, 5%, 1.7%, and 0.2%. The acceptable limit for kVp accuracy is ≤ ± 5. This means that three (3) machines passed the accuracy test and two (2) failed. Further, the computed kVp reproducibility results for all machines are 10%, 10%, 9%, 2.8%, and 1.2%. The allowable limit is ≤ ± 5, indicating that only two of the X-ray machines passed the reproducibility test. Quality assurance in a diagnostic radiological facility ensures that exposure parameters are calibrated, that the X-ray machine's functional performance is adequate, and that radiation safety is maintained within the X-ray room. This is normally done to obtain the optimal diagnostic data at the lowest risk) to the patient. Routine checks are carried out to ensure that the values of the radiation output by the X-ray equipment are within the allowable limit specified by the regulatory authority. Quality assurance tests are performed at regular intervals (once a year) and at major repairs of the X-ray machine, and should be done by a qualified service engineer, diagnostic medical physicist, or a well-trained QC technologist at a facility approved by the regulatory body. The failure of some machines is due to extremely high kVp, which increases the penetration of the beam and affects the image contrast, while a lower kVp leads to inadequate beam penetration and increased image noise. Incorrect kVp leads to inappropriate patient exposure and possibly repeat examinations. Reproducibility shows the highest failure rate because it is sensitive to instability in the X-ray machine. This could be due to such factors as machine age, inadequate maintenance culture, or calibration. Equipment age may be associated with deterioration of generator components, X-ray tubes, exposure-control systems, and electrical components. Inadequate preventive maintenance may accelerate such deterioration. Research conducted on medical X-ray machines in Tanzania revealed that 85.71% and 71.43% passed the reproducibility and mAs linearity tests, respectively. A report from Nigeria indicated that out of the ten (10) X-ray equipment evaluated for kVp accuracy, reproducibility, mAs linearity, and beam alignment tests, six (6) were within the acceptable limits and four (4) failed the quality control parameters. However, there is no evidence of published work that specifically highlights the quality control test on diagnostic X-rays in Freetown, Sierra Leone, that focuses on measuring the kVp accuracy, mAs linearity, and reproducibility. The significance of evaluating the five facilities is based on them serving as a baseline assessment or evidence that can support the development of quality control programs, maintenance planning, regulatory control, and a national X-ray diagnostic survey in the future.
Accuracy evaluates whether the measured output corresponds to the selected value, whereas reproducibility evaluates whether the equipment can repeatedly produce essentially the same output under identical exposure conditions. A failure in kVp reproducibility causes a variation in radiation output and appearance of the image, which may lead to repeat examinations. Poor kVp reproducibility is an indication that identical console settings may produce different X-ray tube voltages on repeated exposures. Poor mAs linearity shows that changes in mAs do not lead to proportional changes in radiation output, hence reducing the prediction of patient dose and image receptor exposure. Failure of quality control parameters leads to non-diagnostic images that require exposure to be repeated, causing additional patient radiation exposure. This also increases the workload of radiologists and radiographers. A consistent quality control failure indicates a deterioration of the X-ray machine, which requires preventive or corrective maintenance.
Unprecedentedly, this research provides a baseline assessment of the performance of quality control parameters of selected diagnostic X-ray machines in Freetown, Sierra Leone, where there is currently no empirical evidence of published work on the quality control performance of X-ray machines. The evaluation of mAs linearity, kVp accuracy, and reproducibility provides initial evidence on the functional performance of diagnostic X-ray equipment, which can be used as a baseline to compare future quality control assessments of medical diagnostic facilities. According to the findings, the failure of quality control parameters of some X-rays affects or has implications on patient dose, image quality, and equipment maintenance. This work is, however, limited to only five selected diagnostic facilities in Freetown, focusing only on three selected parameters, and did not assess other significant quality control parameters such as half-value layer, radiation output, exposure-time accuracy, beam alignment, collimation, image quality and patient entrance-surface dose.
4. Conclusion
Quality control tests were carried out on five (5) different X-ray equipment in selected diagnostic hospitals in Freetown. Three parameters (mAs Linearity, kVp accuracy, and reproducibility) were evaluated to determine whether the radiation dose to patients meets the standard regulatory requirement. The computed figures show that four results for the mAs linearity test are within the tolerance level. Three values for kVp accuracy passed, while two are below the limits. For reproducibility, two X-ray machines passed the test, and three failed. Inconsistency in radiation output may lead to high radiation exposure and poor or degraded image quality. Corrective actions are required for all X-ray machines, with special emphasis on equipment with extremely inadequate/default reproducibility results. Future work will capture a good number of diagnostic X-ray facilities in Freetown and other parts of the country to establish a national baseline for the operational performance of X-ray equipment in Sierra Leone. Further studies should also incorporate additional quality-control parameters such as exposure-time accuracy, half-value layer, beam alignment, collimation, image-quality assessment, etc. The establishment of a national database for diagnostic X-ray QC results would also support regulatory monitoring, preventive maintenance, and evidence-based radiation protection programs.
Abbreviations

PRD

Personal Radiation Detector

mAs

Milliampere seconds

kVp

Kilovoltage

CoL

Coefficient of Linearity

mR

Millroentgen

SD (σ)

Standard Deviation

CV

Coefficient of Variation

µGy

Microgray

mGy

Milligray

Acknowledgments
This work was facilitated by the Faculty of Engineering and Technology, Ernest Bai Koroma University of Science and Technology, and the Nuclear Safety and Radiation Protection Authority.
Author Contributions
Saidu Kamara: Conceptualization, Data curation, Investigation, Methodology, Resources, Writing – original draft, Writing – review & editing
Conflicts of Interest
The author declares no conflicts of interest.
References
[1] A. Barzylovych, Y. Ursakii, A. Nadezhdenko, T. Mamatova, O. Chykarenko, and S. Kravchenko, "The influence of medical services public management on the population’life quality," 2021.
[2] A. Mohammad Mosadeghrad, "Healthcare service quality: towards a broad definition," International journal of health care quality assurance, vol. 26, pp. 203-219, 2013.
[3] S. Code, "Safety Procedures for the Installation, Use and Control of X-ray Equipment in Large Medical Radiological Facilities," 2024.
[4] M. Amurao, D. A. Gress, M. A. Keenan, P. H. Halvorsen, J. A. Nye, and M. Mahesh, "Quality management, quality assurance, and quality control in medical physics," Journal of applied clinical medical physics, vol. 24, p. e13885, 2023.
[5] H. Delis, K. Christaki, B. Healy, G. Loreti, G. L. Poli, P. Toroi, et al., "Moving beyond quality control in diagnostic radiology and the role of the clinically qualified medical physicist," Physica Medica, vol. 41, pp. 104-108, 2017.
[6] G. K. Korir, J. S. Wambani, and I. K. Korir, "Establishing a quality assurance baseline for radiological protection of patients undergoing diagnostic radiology," SA Journal of Radiology, vol. 15, pp. 70-79, 2011.
[7] L. L. Nkuba and P. B. Nyanda, "Compliance and Quality Control Monitoring of Diagnostic X ray Facilities in Dar es Salaam City, Tanzania," Brazilian Journal of Radiation Sciences, vol. 5, 2017.
[8] M. A. Ali, S. A. Vuai, S. A. Suleiman, and M. F. Mwema, "Assessment of regulatory compliance in diagnostic X-ray facilities based on a radiation safety audit in Zanzibar," Journal of Medical Imaging and Radiation Sciences, vol. 57, p. 102190, 2026.
[9] J. Alyami and M. Nassef, "Assessment of diagnostic radiology facilities technical radiation protection requirements in KSA," Applied Sciences, vol. 12, p. 7284, 2022.
[10] B. Van der Merwe, M. Nel, and S. Kruger, "Radiation safety requirements for training of users of diagnostic X-ray equipment in South Africa," African Journal of Health Professions Education, vol. 9, pp. 123-127, 2017.
[11] A. Sonawane, M. Singh, J. S. Kumar, A. Kulkarni, V. Shirva, and A. Pradhan, "Radiological safety status and quality assurance audit of medical X-ray diagnostic installations in India," Journal of Medical Physics, vol. 35, pp. 229-234, 2010.
[12] M. Sandborg, J. N. Althén, and A. Gustafsson, "Efficient quality assurance programs in radiology and nuclear medicine in Östergötland, Sweden," Radiation protection dosimetry, vol. 139, pp. 410-417, 2010.
[13] V. Patel, G. Sindhwani, M. Gupta, S. Arora, A. Mishra, J. Bhatt, et al., "A comprehensive approach towards quality and safety in diagnostic imaging services: our experience at a rural tertiary health care center," Journal of Clinical and Diagnostic Research: JCDR, vol. 11, p. TC10, 2017.
[14] B. Allen, M. Chatfield, J. Burleson, and W. T. Thorwarth, "Improving diagnosis in health care: perspectives from the American College of Radiology," Diagnosis, vol. 4, pp. 113-124, 2017.
[15] M. Lötter, "Development and evaluation of a quality assurance programme for a radiology department," 2018.
[16] J. C. Broder, S. F. Cameron, W. T. Korn, and S. J. Baccei, "Creating a radiology quality and safety program: principles and pitfalls," Radiographics, vol. 38, pp. 1786-1798, 2018.
[17] I. P. 154, "Optimisation of radiological protection in digital radiology techniques for medical imaging," Annals of the ICRP, vol. 52, pp. 11-145, 2023.
[18] C. Cloete, "An activity-based workload modelling approach for determining diagnostic radiographer staffing requirements within a diagnostic radiology practice," Stellenbosch: Stellenbosch University, 2021.
[19] D. Polevoy, D. Kazimirov, M. Gilmanov, and D. Nikolaev, "No reproducibility, No progress: rethinking CT benchmarking," Journal of Imaging, vol. 11, p. 344, 2025.
[20] T. Chusin, R. Wongchai, S. Moonkham, T. Pengpan, and K. Aphiwatthanasumet, "Evaluation of the reproducibility of automatic exposure control systems in general X-ray machines using a coin-based method: " Radiological Physics and Technology, vol. 18, pp. 1239-1246, 2025.
[21] N. I. Barau, W. E. Mangset, I.-O. M. Ijeoma, and Y. B. Mohammed, "Evaluation of Quality Control Standards in Medical Diagnostic X-Ray Facilities in North-Eastern Nigeria," Nigerian Journal of Physics, vol. 34, pp. 110-117, 2025.
[22] A. Pilipović, G. Baršić, M. Katić, and M. Rujnić Havstad, "Repeatability and reproducibility assessment of a PolyJet technology using x-ray computed tomography," Applied Sciences, vol. 10, p. 7040, 2020.
[23] D. Prakash and R. P. Kotian, "Standard X-Ray Tubes: Basic Principles, Types of X-Ray Tubes, and Routine Quality Control," in Fundamentals of X-ray Imaging: Basic Principles, Quality Control, Clinical Applications, and Safety, ed: Springer, 2025, pp. 87-184.
[24] M. Anburajan and J. K. Sharma, "“Overview of X-Ray Tube Technology” 21," Biomedical engineering and its applications in healthcare, p. 519, 2019.
[25] D. Meikle and G. Hill, "X‐ray Production," Practical Radiotherapy: Physics and Equipment, 3rd Edition, pp. 23-33, 2019.
[26] R. Pauwels, O. Silkosessak, R. Jacobs, R. Bogaerts, H. Bosmans, and S. Panmekiate, "A pragmatic approach to determine the optimal kVp in cone beam CT: balancing contrast-to-noise ratio and radiation dose," Dentomaxillofacial Radiology, vol. 43, p. 20140059, 2014.
[27] E. Seeram, R. Davidson, S. Bushong, and H. Swan, "Radiation dose optimization research: Exposure technique approaches in CR imaging-A literature review," Radiography, vol. 19, pp. 331-338, 2013.
[28] G. X. Ding and P. Munro, "Radiation exposure to patients from image guidance procedures and techniques to reduce the imaging dose," Radiotherapy and Oncology, vol. 108, pp. 91-98, 2013.
[29] C. R. Kim and J. Y. Jeon, "Radiation dose and image conspicuity comparison between conventional 120 kVp and 150 kVp with spectral beam shaping for temporal bone CT," European Journal of Radiology, vol. 102, pp. 68-73, 2018.
[30] A. A. Oglat, "Comparison of X-ray films in term of kVp, mA, exposure time and distance using Radiographic Chest Phantom as a radiation quality," Journal of Radiation Research and Applied Sciences, vol. 15, p. 100479, 2022.
[31] T. L. Fauber, T. F. Cohen, and M. C. Dempsey, "High kilovoltage digital exposure techniques and patient dosimetry," Radiologic technology, vol. 82, 2011.
[32] A. Almutairi and A. A. Mohammed, "Performance assessment of GE conventional X-ray System installed in Radiological Science Department of King Saud University," Am. J. PharmTcch. Res, vol. 14, 2024.
[33] M. Hashemi, S. Bayani Roodi, F. Shahedi, M. Momennezhad, H. Zare, and H. Gholamhosseinian, "Quality assessment of conventional X-ray diagnostic equipment by measuring X-ray exposure and tube output parameters in Great Khorasan Province, Iran," Iranian Journal of Medical Physics, vol. 16, pp. 34-40, 2019.
[34] R. A. Lynch, T. Smith, M. C. Jacobs, S. J. Frank, and K. J. Kearfott, "A radiation weather station: development of a continuous monitoring system for the collection, analysis, and display of environmental radiation data," Health Physics, vol. 115, pp. 590-599, 2018.
[35] S. Pommé, J. Paepen, and R. Van Ammel, "Linearity check of an ionisation chamber through 99 mTc half-life measurements," Applied Radiation and Isotopes, vol. 140, pp. 171-178, 2018.
[36] J. P. Bissonnette, D. J. Moseley, and D. A. Jaffray, "A quality assurance program for image quality of cone‐beam CT guidance in radiation therapy," Medical physics, vol. 35, pp. 1807-1815, 2008.
[37] F. Zarb, L. Rainford, and M. F. McEntee, "Image quality assessment tools for optimization of CT images," Radiography, vol. 16, pp. 147-153, 2010.
[38] H. Herath, H. Herath, N. Madusanka, and B.-I. Lee, "A systematic review of medical image quality assessment," Journal of Imaging, vol. 11, p. 100, 2025.
[39] B. Ijabor, C. Nzotta, and A. Omojola, "Quality control test of conventional X-Ray systems in Delta State, South-South, Nigeria," World Scientific News, vol. 157, pp. 140-153, 2021.
[40] M. P. Kadima, "Assessment of quality control in medical diagnostic x-ray facilities in the western region of Kenya," 2013.
[41] U. Hirvonen, "X-Ray Compliance Testing of a Mammography System on Manufacturing Line: Dose Reproducibility & Linearity," 2015.
[42] P. T. Hornsby, "Simplified Quality Control System in Contrast to Performance Measurements to Assure Minimized Patient Exposure From Diagnostic X-Ray Machines," Idaho State University, 2024.
[43] I. I. Suliman, "Patient dosimetry and quality control in diagnostic radiology," Sudan Academy of Sciences, Khartoum (Sudan)2007.
[44] E. Karavas, B. Ece, S. Aydın, M. Kocak, Z. Cosgun, I. E. Bostanci, et al., "Are we aware of radiation: A study about necessity of diagnostic X-ray exposure," World Journal of Methodology, vol. 12, p. 264, 2022.
[45] E. A. Alakrout, M. M. Zayed, A. A. Al-Lisini, and H. A. Al-Tawerghi, "Public Awareness and Knowledge of Ionizing Radiation Risks from Medical Imaging Examinations in Western Libya," Surman Journal of Science and Technology, vol. 7, pp. 105-117, 2025.
[46] J. Ford, "Radiation, people and the environment," 2004.
[47] S. Visweswaran, K. Kanagaraj, S. Joseph, and V. Perumal, "Medical imaging: Contribution toward background radiation and human exposure," Journal of Radiation and Cancer Research, vol. 9, pp. 177-182, 2018.
[48] A. R. A. Mishal, M. S. Jabr, and Z. I. Muhammad, "Radiation: Types, Natural sources, and Industrial Sources," Current Clinical and Medical Education, vol. 2, pp. 89-94, 2024.
[49] V. Harahap, J. Saragih, and E. B. Ginting, "IMPLEMENTATION OF RADIATION PROTECTION QUALITY CONTROL IN RADIOLOGY INSTALLATIONS IN ACCORDANCE WITH MINISTRY OF HEALTH REGULATION NO. 8 OF 2011," Jurnal Ilmiah METADATA, vol. 5, pp. 456-478, 2023.
[50] G. Fan and Q. Wang, "Quality control and quality assurance," in Clinical molecular diagnostics, ed: Springer, 2021, pp. 97-113.
[51] H. H. van Rossum, "Technical quality assurance and quality control for medical laboratories: a review and proposal of a new concept to obtain integrated and validated QA/QC plans," Critical reviews in clinical laboratory sciences, vol. 59, pp. 586-600, 2022.
Cite This Article
  • APA Style

    Kamara, S. (2026). Evaluation of mAs Linearity, kVp Accuracy, and Reproducibility in Selected X-ray Diagnostic Facilities in Freetown. Radiation Science and Technology, 12(2), 31-43. https://doi.org/10.11648/j.rst.20261202.12

    Copy | Download

    ACS Style

    Kamara, S. Evaluation of mAs Linearity, kVp Accuracy, and Reproducibility in Selected X-ray Diagnostic Facilities in Freetown. Radiat. Sci. Technol. 2026, 12(2), 31-43. doi: 10.11648/j.rst.20261202.12

    Copy | Download

    AMA Style

    Kamara S. Evaluation of mAs Linearity, kVp Accuracy, and Reproducibility in Selected X-ray Diagnostic Facilities in Freetown. Radiat Sci Technol. 2026;12(2):31-43. doi: 10.11648/j.rst.20261202.12

    Copy | Download

  • @article{10.11648/j.rst.20261202.12,
      author = {Saidu Kamara},
      title = {Evaluation of mAs Linearity, kVp Accuracy, and Reproducibility in Selected X-ray Diagnostic Facilities in Freetown},
      journal = {Radiation Science and Technology},
      volume = {12},
      number = {2},
      pages = {31-43},
      doi = {10.11648/j.rst.20261202.12},
      url = {https://doi.org/10.11648/j.rst.20261202.12},
      eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.rst.20261202.12},
      abstract = {Quality control in medical radiology is a process used by a regulatory authority to evaluate the performance of a diagnostic X-ray machine, including acceptable limits and the actions necessary to meet regulatory requirements. This study evaluates the radiation dose outputs of X-ray equipment in diagnostic medical hospitals by measuring mAs linearity, kVp accuracy, and reproducibility as quality control parameters to ensure that radiation doses administered to patients meet standard regulatory requirements. Radiation measuring instruments were used in this work to measure mAs, kVp accuracy, and reproducibility outputs for five (5) medical diagnostic X-ray machines. The detector was placed on the couch at a distance of 100 cm from the collimator to take measurements for each parameter under investigation. Dose rate measurements for each parameter were recorded one at a time. The mathematical equations for mAs linearity, kVp accuracy, and reproducibility were used to compute the linearity coefficient, percentage error, and coefficient of variation. The computed figures for mAs linearity for the five X-ray machines show that four (4) passed and one fell below the allowable limit. Data for kVp accuracy indicate that three of the machines are within the allowable limits and two (2) failed, while the calculated values for reproducibility show that only two (2) machines passed the tests. The study generates empirical quality control data rarely reported in X-ray diagnostic centers. It verifies the linearity coefficient, percentage kVp error, and the coefficient of variation with practical knowledge in radiology studies. Additionally, it provides a systematic mathematical approach to compute mAs linearity, kVp accuracy, and reproducibility for teaching radiology in non-diagnostic laboratory settings.},
     year = {2026}
    }
    

    Copy | Download

  • TY  - JOUR
    T1  - Evaluation of mAs Linearity, kVp Accuracy, and Reproducibility in Selected X-ray Diagnostic Facilities in Freetown
    AU  - Saidu Kamara
    Y1  - 2026/10/21
    PY  - 2026
    N1  - https://doi.org/10.11648/j.rst.20261202.12
    DO  - 10.11648/j.rst.20261202.12
    T2  - Radiation Science and Technology
    JF  - Radiation Science and Technology
    JO  - Radiation Science and Technology
    SP  - 31
    EP  - 43
    PB  - Science Publishing Group
    SN  - 2575-5943
    UR  - https://doi.org/10.11648/j.rst.20261202.12
    AB  - Quality control in medical radiology is a process used by a regulatory authority to evaluate the performance of a diagnostic X-ray machine, including acceptable limits and the actions necessary to meet regulatory requirements. This study evaluates the radiation dose outputs of X-ray equipment in diagnostic medical hospitals by measuring mAs linearity, kVp accuracy, and reproducibility as quality control parameters to ensure that radiation doses administered to patients meet standard regulatory requirements. Radiation measuring instruments were used in this work to measure mAs, kVp accuracy, and reproducibility outputs for five (5) medical diagnostic X-ray machines. The detector was placed on the couch at a distance of 100 cm from the collimator to take measurements for each parameter under investigation. Dose rate measurements for each parameter were recorded one at a time. The mathematical equations for mAs linearity, kVp accuracy, and reproducibility were used to compute the linearity coefficient, percentage error, and coefficient of variation. The computed figures for mAs linearity for the five X-ray machines show that four (4) passed and one fell below the allowable limit. Data for kVp accuracy indicate that three of the machines are within the allowable limits and two (2) failed, while the calculated values for reproducibility show that only two (2) machines passed the tests. The study generates empirical quality control data rarely reported in X-ray diagnostic centers. It verifies the linearity coefficient, percentage kVp error, and the coefficient of variation with practical knowledge in radiology studies. Additionally, it provides a systematic mathematical approach to compute mAs linearity, kVp accuracy, and reproducibility for teaching radiology in non-diagnostic laboratory settings.
    VL  - 12
    IS  - 2
    ER  - 

    Copy | Download

Author Information
  • Faculty of Engineering, Ernest Bai Koroma University of Science and Technology, Magburaka, Sierra Leone;Nuclear Safety and Radiation Protection Authority, Ministry of Environment, Freetown, Sierra Leone